Fluke MET/CAL Procedure ============================================================================= INSTRUMENT: Fluke 16: (1 year) CAL VER /5520 DATE: 22-Sep-98 AUTHOR: Fluke Corporation REVISION: $Revision: 1.2 $ ADJUSTMENT THRESHOLD: 70% NUMBER OF TESTS: 50 NUMBER OF LINES: 217 CONFIGURATION: Fluke 5520A ============================================================================= # # Source: # Fluke-16 and Fluke-16INTL Service Information Package For Service Centers # (PN 644226, June 1997) # # Compatibility: # 5500/CAL 5.0 or MET/CAL 5.0 or later # # Subprocedures: # None # # Required Files: # 52_18_2w.bmp # 52_16_la.bmp # 52_16_k.bmp # # System Specifications: # TUR calculation is based on specification interval of the accuracy file. # The default 5520A accuracy file contains 90 day specs. # # Fluke makes no warranty, expressed or implied, as to the fitness # or suitability of this procedure in the customer's application. # STEP FSC RANGE NOMINAL TOLERANCE MOD1 MOD2 3 4 CON 1.001 ASK- R N P F V 1.002 ASK+ K 1.003 HEAD EQUIPMENT SETUP 1.004 DISP [32] WARNING 1.004 DISP HIGH VOLTAGE is used or exposed during the performance 1.004 DISP of this calibration. DEATH ON CONTACT may result if 1.004 DISP personnel fail to observe safety precautions. 1.005 DISP No warm-up time requirement. 1.005 DISP Ambient temperature: 18C - 28C. 1.005 DISP Relative humidity: less than 90%. 1.006 DISP Set the Rotary Switch to the OFF position. 1.006 DISP 1.006 DISP Note: Erroneous result may occur if switch is not 1.006 DISP placed in OFF position before beginning tests. 1.007 HEAD {AC VOLTAGE TESTS} 1.008 PIC 52_18_2W 1.009 DISP Set the Rotary Switch to the center position. 1.010 HEAD AC VOLTAGE TESTS: {4V Range (HiZ)} 1.011 DISP Press RANGE to select the 4V range (MAN 0.000). 1.012 5520 0.000V S 2W 1.013 MEMI Enter UUT reading in volts: 1.014 MEME 1.015 MEMC 4 V +0.003U 2.001 ASK- Q 2.002 5520 0.000V S 2W 2.003 HEAD AC VOLTAGE TESTS: {4V Range (MAX)} 2.004 DISP Press MAX/MIN to select Min/Max Mode. 2.004 DISP Press MAX/MIN again to display Max. 2.005 MEMI Enter UUT reading in volts: 2.006 MEME 2.007 MEMC 4 V +0.040U 3.001 MEME 3.002 HEAD AC VOLTAGE TESTS: {4V Range (MIN)} 3.003 DISP Press MAX/MIN to display Min. 3.004 MEMI Enter UUT reading in volts: 3.005 MEME # Lower limit should be 0.000 but firmware bug in early units allowed -0.004 3.006 MEMC 4 V -.004U +.040U 4.001 DISP Press MAX/MIN for 2s to exit Min/Max Mode. 4.002 ASK+ Q 4.003 HEAD AC VOLTAGE TESTS: {4V Range (HiZ)} 4.004 5520 4 0.010V 0.003U 50H SI 2W 5.001 5520 4 0.010V 0.003U 400H SI 2W 6.001 HEAD AC VOLTAGE TESTS: {4V Range (LoZ)} 6.002 DISP Set the Rotary Switch to the right position. 6.003 5520 4 3.500V 0.070U 50H SI 2W 7.001 HEAD AC VOLTAGE TESTS: {4V Range (HiZ)} 7.002 DISP Set the Rotary Switch to the center position. 7.003 5520 4 3.500V 0.070U 400H SI 2W 8.001 HEAD AC VOLTAGE TESTS: {40V Range (HiZ)} 8.002 DISP Press RANGE to select the 40V range (MAN 0.00). 8.003 5520 40 35.00V 0.70U 50H SI 2W 9.001 DISP Set the Rotary Switch to the right position. 9.002 HEAD AC VOLTAGE TESTS: {40V Range (LoZ)} 9.003 5520 40 35.00V 0.70U 400H SI 2W 10.001 HEAD AC VOLTAGE TESTS: {400V Range (LoZ)} 10.002 DISP Press RANGE to select the 400V range (MAN 0.0). # Warmup 2k resistor so 350V will not current limit 5520A. 10.003 5520 100.0V 80H SI S 2W 10.004 PORT [D1000] 10.005 5520 150.0V 80H SI S 2W 10.006 PORT [D1000] 10.007 5520 200.0V 80H SI S 2W 10.008 PORT [D1000] 10.009 5520 250.0V 80H SI S 2W 10.010 PORT [D1000] 10.011 5520 300.0V 80H SI S 2W 10.012 PORT [D1000] 10.013 5520 400 350.0V 7.0U 80H SI 2W 11.001 HEAD AC VOLTAGE TESTS: {600V Range (HiZ)} 11.002 DISP Set the Rotary Switch to the center position. 11.002 DISP Press RANGE to select the 600V Range (MAN 0). 11.003 5520 600 600V 14U 400H SI 2W 12.001 HEAD {DC VOLTAGE TESTS}: 4V Range 12.002 DISP Press SELECT for VDC. 12.002 DISP Press RANGE to select the 4V Range (MAN 0.000). 12.003 ASK- Q 12.004 5520 0.000V S 2W 12.005 HEAD DC VOLTAGE TESTS: {4V Range (MAX)} 12.006 DISP Press MAX/MIN to select Min/Max Mode. 12.006 DISP Press MAX/MIN again to display Max. 12.007 MEMI Enter UUT reading in volts: 12.008 MEME 12.009 MEMC 4 V 0.012U 13.001 MEME 13.002 HEAD DC VOLTAGE TESTS: {4V Range (MIN)} 13.003 DISP Press MAX/MIN to display Min. 13.004 MEMI Enter UUT reading in volts: 13.005 MEME 13.006 MEMC 4 V 0.012U 14.001 DISP Press MAX/MIN for 2s to exit Min/Max Mode. 14.002 ASK+ Q 14.003 HEAD DC VOLTAGE TESTS: {4V Range (LoZ)} 14.004 DISP Set the Rotary Switch to the right position. 14.005 5520 4 0.000V 0.002U 2W 15.001 HEAD DC VOLTAGE TESTS: {40V Range (LoZ)} 15.002 DISP Press RANGE to select the 40V Range (MAN 0.00). 15.003 5520 40 0.02V 0.01U 2W 16.001 HEAD DC VOLTAGE TESTS: {400V Range (HiZ)} 16.002 DISP Set the Rotary Switch to the center position. 16.003 DISP Press RANGE to select the 400V Range (MAN 0.0). 16.004 5520 400 -0.2V 0.1U 2W 17.001 HEAD DC VOLTAGE TESTS: {4V Range (HiZ)} 17.002 DISP Press RANGE for 2s to select autorange. 17.003 5520 4 3.500V 0.034U 2W 18.001 HEAD DC VOLTAGE TESTS: {40V Range (HiZ)} 18.002 5520 40 -35.00V 0.33U 2W 19.001 HEAD DC VOLTAGE TESTS: {400V Range (HiZ)} 19.002 5520 400 100.0V 1.0U 2W 20.001 HEAD DC VOLTAGE TESTS: {600V Range (HiZ)} 20.002 5520 600 -600V 6U 2W 21.001 HEAD {RESISTANCE TESTS} 21.002 DISP Set the Rotary Switch to the right position. 21.003 HEAD RESISTANCE TESTS: {400 Ohm Range} 21.004 5520 400 0.0Z +0.2U 2W 22.001 5520 400 1.0Z 0.2U 2W 23.001 5520 400 350.0Z 3.4U 2W 24.001 HEAD RESISTANCE TESTS: {4 kOhm Range} 24.002 5520 4 1.000kZ 0.010U 2W 25.001 HEAD RESISTANCE TESTS: {40 kOhm Range} 25.002 5520 40 35.00kZ 0.33U 2W 26.001 HEAD RESISTANCE TESTS: {400 kOhm Range} 26.002 5520 400 100.0kZ 1.0U 2W 27.001 HEAD RESISTANCE TESTS: {4 MOhm Range} 27.002 5520 4 2.900MZ 0.027U 2W 28.001 HEAD RESISTANCE TESTS: {40 MOhm Range} 28.002 5520 40 35.00MZ 0.56U 2W 29.001 HEAD RESISTANCE TESTS: {VCHEK} 29.002 5520 1.960V 0.9% 0.002U 2W 30.001 HEAD {CONTINUITY TEST} 30.002 DISP Press SELECT for Continuity. 30.003 5520 2.000V 0.020U 2W 31.001 HEAD CONTINUITY TEST: {VCHEK} 31.002 5520 3.430V 0.033U 2W 32.001 5520 -0.500V 0.007U 2W 33.001 HEAD {CAPACITANCE TESTS} 33.002 DISP Press SELECT for Capacitance. 33.003 HEAD CAPACITANCE TESTS: {1uF Range} 33.004 5520 1 0.950uF 0.020U 2W 34.001 DISP Disconnect the 5520A from the UUT. 34.002 MATH MEM1 = 0.0 34.003 MEMI Enter UUT reading in microfarads: 34.004 MEME 34.005 MEMC 1 uF 0.001U 35.001 HEAD {DC CURRENT TESTS}: 200uA Range 35.002 DISP Press SELECT for uA DC. 35.003 PIC 52_16_la 35.004 HEAD DC CURRENT TESTS: {200uA Range} 35.005 5520 200 0.0uA 0.2U 2W 36.001 5520 200 10.0uA 0.3U 2W 37.001 5520 200 -10.0uA 0.3U 2W 38.001 5520 200 200.0uA 2.2U 2W 39.001 5520 200 -200.0uA 2.2U 2W 40.001 HEAD {AC CURRENT TESTS}: 200uA Range 40.002 DISP Press RANGE for uA AC. 40.003 HEAD AC CURRENT TESTS: {200uA Range} # 29uA used in lieu of 20uA. 40.004 5520 200 29.0uA 2% 0.3U 50H SI 2W 41.001 5520 200 29.0uA 2% 0.3U 400H SI 2W 42.001 5520 200 200.0uA 4.3U 50H SI 2W 43.001 5520 200 200.0uA 4.3U 400H SI 2W 44.001 HEAD {TEMPERATURE} 44.002 DISP Press TEMP degC/degF for Temperature (degF). 44.003 PIC 52_16_k 44.004 5520 32.0degF 1.5U _K TC 2W 45.001 5520 73.4degF 2.2U _K TC 2W 46.001 5520 212.0degF 3.6U _K TC 2W 47.001 DISP Press TEMP degC/degF for Temperature (degC). 47.002 5520 0.0degC 0.8U _K TC 2W 48.001 5520 23.0degC 1.0U _K TC 2W 49.001 5520 100.0degC 1.8U _K TC 2W 50.001 DISP Set the Rotary Switch to the OFF position. 50.002 END